Abstract

In this study, albedo parameters are experimentally measured to determine radiation scattering features of some thermoluminescent dosimetry materials (TLD 100, TLD 100h, TLD 200 and TLD 400). In experimental setup, it was used scattering angle of 168°, 59.54 keV gamma rays and HPGe detector. The effective atomic numbers of the investigated materials were determined from the coherent to Compton scattering intensity ratio. Also, the mass attenuation coefficient, the mass energy absorption coefficient and kerma relative to air and ICRU tissue were calculated for the TLD materials. It was observed that TLD 100h is the best tissue equivalent material among other TLD materials and best reflects 59.54 keV gamma rays.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.