Abstract

To compare a 3D preoperative planning image and fluoroscopic image, a 3D bone position estimation system that displays 3D images in response to changes in the position of fluoroscopic images was developed. The objective of the present study was to evaluate the accuracy of the estimated position of 3D bone images with reference to fluoroscopic images. Bone positions were estimated from reference points on a fluoroscopic image compared with those on a 3D image. The four reference markers positional relationships on the fluoroscopic image were compared with those on the 3D image to evaluate whether a 3D image may be drawn by tracking positional changes in the radius model. Intra-class correlations coefficients for reference marker distances between the fluoroscopic image and 3D image were 0.98–0.99. Average differences between measured values on the fluoroscopic image and 3D bone image for each marker corresponding to the direction of the bone model were 1.1 ± 0.7 mm, 2.4 ± 1.8 mm, 1.4 ± 0.8 mm, and 2.0 ± 1.6 mm in the anterior-posterior view, ulnar side lateral view, posterior-anterior view, and radial side lateral view, respectively. Marker positions were more accurate in the anterior-posterior and posterior-anterior views than in the radial and ulnar side lateral views. This system helps in real-time comparison of dynamic changes in preoperative 3D and intraoperative fluoroscopy images.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call