Abstract

Streaming is shown to occur in water in the focused beams produced by a number of medical pulseecho devices. The use of hot film anemometry to measure the streaming velocity is described and velocities measured in water using commercial equipment are quoted. The highest velocities occur in pulsed Doppler mode with a maximum velocity of 14 cm s−1 being observed. An experimental set-up was used to investigate the parameters affecting streaming and it was found that the harmonic content of the pulse waveform had a major effect on the streaming velocity. The time taken for a stream to become established at the focus of the acoustic beams studied was typically approximately 0.5 s.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.