Abstract

Backgroundthe chemometric processing of second-order chromatographic-spectral data is usually carried out with the aid of multivariate curve resolution-alternating least-squares (MCR-ALS). Recently, an alternative procedure was described based on the estimation of image moments for each data matrix and subsequent application of multiple linear regression after suitable variable selection. ResultsThe analysis of both simulated and experimental data leads to the conclusion that the image moment method, although can cope with chromatographic lack of reproducibility across injections, it only performs well in the absence of uncalibrated interferents. MCR-ALS, on the other hand, provides good analytical results in all studied situations, whether the test samples contain uncalibrated interferents or not. SignificanceThe results are useful to assess the real usefulness of newly proposed methodologies for second-order calibration in the case of chromatographic-spectral data sets, especially when samples contain unexpected chemical constituents.