Abstract
The usability of an ex vivo human lung model for ablation of lung cancer tissue with high-intensity focused ultrasound (HIFU) is described. Lung lobes were flooded with saline, with no gas remaining after complete atelectasis. The tumor was delineated sono-morphologically. Speed of sound, tissue density and ultrasound attenuation were measured for flooded lung and different pulmonary cancer tissues. The acoustic impedance of lung cancer tissue (1.6–1.9 mega-Rayleighs) was higher than that of water, as was its attenuation coefficient (0.31–0.44 dB/cm/MHz) compared with that of the flooded lung (0.12 dB/cm/MHz). After application of HIFU, the temperature in centrally located lung cancer surrounded by the flooded lung increased as high as 80°C, which is sufficient for treatment. On the basis of these preliminary results, ultrasound-guided HIFU ablation of lung cancer, by lung flooding with saline, appears feasible and should be explored in future clinical studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have