Abstract

Three new immunoassays for determination of serum gentamicin concentration were assessed: a latex agglutination assay (Technicon), an enzyme immunoassay (EMIT 2000) and a multi-layer thin-film immunoassay (Opus). All three were easy to use, had good within- and between-assay reproducibility, good reliability and acceptable linearity. The standard curve stability was shortest with the latex agglutination assay and longest with the thin-film. The thin-film assay calibrators gave approximately double the expected concentration when assayed by polarisation fluoroimmunoassay but this did not effect the assay's performance. When compared with the polarisation fluoroimmunoassay results of patient samples, all three methods gave an excellent correlation. The technical aspects of these three methods make them acceptable for clinical use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.