The TruView EVO2 laryngoscope was compared with the traditional Macintosh laryngoscope in 200 patients who required tracheal intubation for elective surgery. Mallampati score determined prior to laryngoscopy was significantly related to the view of the glottis during laryngoscopy for both laryngoscopes. The view of the larynx was better with the TruView EVO2 laryngoscope than with the Macintosh laryngoscope in patients with a Cormack and Lehane grade greater than 1 (p < 0.01). The mean time to intubate was significantly shorter with the Macintosh laryngoscope (34 s) than with the TruView laryngoscope (51 s) (p < 0.01).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.