Abstract
We evaluated the manufacturer's temperature calibration procedure for the Union Carbide Centrifichem 400 using a procedure developed in our laboratories. We substituted a Yellow Springs, Inc., Model 45 CU Cuvette Thermometer with a thermistor probe for the mercury thermometers. This dramatically shortened the time required for temperature measurements by eliminating the time required for the mercury thermometers to equilibrate. The cuvette temperature measurements with the YSI 45 CU accurately reflect the analyzer equilibrium temperature only when the YSI 45 CU thermistor probe is in the same cuvette as the analyzer thermistor. We also evaluated the accuracy and precision of the temperature control system and the time required to reach thermal equilibrium. We found that temperature equilibration occurs in less than a minute and has a range of less than ±0.1°C from the mean temperature. Our procedure for monitoring temperature equilibration gave improved accuracy and precision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.