Abstract
We describe an evaluation of four ELISA methods, including three commercial kits, for measuring recombinant and natural human interferon- g (hIFN- g). Using a panel of samples, including well-characterized reference standards, we compared relative quantification between assays, within assays and, where possible, the absolute accuracy of quantification as compared to other analytical methods. The four assays generated markedly different results; up to an almost 60-fold difference between the highest and lowest values for one sample. The differences between assays were not necessarily predictable. No single correction factor could be determined to correct results from one method to another across the panel of samples tested. We conclude that investigators should be diligent to revalidate commercial methods before depending on such methods and resultant data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.