Abstract

The following suture materials have been evaluated for their suitability for use in flexor tendon repairs: 4/0 gauge monofilament and multifilament stainless steel, mono-filament nylon, monofilament polypropylene, monofilament polybutestor, braided polyester, braided polyglycolic acid and a monofilament polyglyconate. These were investigated for their tensile strength (both knotted and unknotted), their extension to failure and knot-holding properties. Stainless steel and monofilament polyglyconate appeared to be the most suitable in that they had high tensile strengths, both knotted and unknotted, and had good knot-holding security. The only disadvantages are that stainless steel is difficult to use and monofilament polyglyconate is absorbable. Polypropylene and braided polyester, although having lower tensile strengths, are reasonable alternatives.