Abstract

We investigated cine-mode portal imaging on a Varian Trilogy accelerator and found that the linearity and other dosimetric properties are sufficient for 3D dose reconstruction as used in patient-specific quality assurance for VMAT (RapidArc) treatments. We also evaluated the gantry angle label in the portal image file header as a surrogate for the true imaged angle. The precision is only just adequate for the 3D evaluation method chosen, as discrepancies of 2° were observed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.