Abstract

Membrane plasma separators are being used routinely for therapy in various diseases. In this study, a newly developed plasma separator made of polyethersulfone (PES) hollow fibers was evaluated for its plasma filtration efficiency and blood compatibility by animal experiment. Hemolysis did not occur under the usual conditions of plasma separation. The sieving coefficients of total protein and albumin were over 95%, and the total cholesterol was over 90% throughout the perfusions. Decreases in white blood cells, platelets, fibrinogen, and coagulation factors were observed during the early stage of plasma separation, but appear to be within acceptable ranges for clinical use.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.