Abstract

In conventional radiography systems, it is apparent that only the area immediately around the central x-ray beam can be evaluated accurately. Consequently in some instances, spinal radiography for example, several exposures are needed at various points along the body to create an accurate image for diagnosis. However, if the film and body part are in a concave shape such that the radius of the curve is equal to the film focal distance, the x-ray beam will penetrate the body and strike the film at two-dimensionally right angles in all areas. Using the spine as an example we found the curved technique had three major advantages over the traditional flat technique: lack of distortion, more uniform beam intensity due to a constant focal film distance, and improved resolution at the periphery of the radiograph because of lack of a cross over effect. It was concluded that an accurate evaluation of larger body parts can be made with minimal distortion utilizing the principles of a curved table technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.