Abstract

Handheld indirect calorimetry has the potential to allow simple and inexpensive measurement of resting metabolic rate in spontaneously breathing people. However, validation work on these devices is contradictory. The purpose of the current study was to determine the bias and level of agreement of oxygen consumption and resting metabolic rate as measured by a handheld indirect calorimeter against a standard open-circuit indirect calorimetry cart. One hundred community-living, spontaneously breathing, ambulatory nonobese and obese adults were studied in single sessions by a single investigator. Sequential measurements were undertaken using the handheld indirect calorimeter and the standard metabolic cart. Measurement sequence was varied randomly. The mean value for oxygen consumption and metabolic rate of the 2 devices was not significantly different. However, agreement between the 2 devices was only 43% in nonobese and obese participants, and there was proportional and fixed bias, with the handheld calorimeter tending to produce a higher value for oxygen consumption and resting metabolic rate. Limits of agreement for resting metabolic rate between the 2 calorimeters were -240 to +300 kcal/d. Measurements of resting metabolic rate by the handheld indirect calorimeter tested in this study are not equivalent to measurements by standard indirect calorimetry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.