Abstract
Enzymic measurement of high-density lipoprotein cholesterol with a centrifugal analyzer is described. We used polyethylene glycol (Mr 6000), final concentration 100 g/L, to precipitate low-density and very-low-density lipoproteins, thereby eliminating the difficulties of the commonly used heparin/Mn2+ precipitation method and facilitating the use of ethylenediaminetetraacetate-stabilized plasma. As measured by rocket immunoelectrophoresis, this final concentration of polyethylene glycol completely precipitates beta-lipoproteins, leaving the alpha-lipoproteins in solution. Between-run reproducibility (CV) was 3.6%, within-run reproducibility (CV) 0.8%. Reagent costs currently are $US 0.13 per test and large numbers of samples can be handled conveniently. Normal ranges were compiled for 539 men and 444 women. The high-density lipoprotein cholesterol for men was 1.20 +/- 0.31 (SD) mmol/L and for women 1.52 +/- 0.38 (SD) mmol/L.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.