Abstract

An inexpensive test chamber has been constructed that provides an environment appropriate for testing the effects of temperature and chemical interactions on gaseous emissions from test solutions. Temperature, relative humidity, and ventilation rate can be controlled and a well-mixed atmosphere can be maintained. The system is relatively simple and relies on heated tap water or ice to adjust the temperature. Temperatures ranging from 9 to 21 degrees C have been maintained. At an average temperature of 15.1 degrees C, temperatures at any location within the chamber vary by no more than 0.5 degree C, and the temperature of the test solution within the chamber varies by no more than 0.1 degree C. The temperatures within the chamber are stable enough to generate precise steady-state concentrations. The wind velocities within the chamber are reproducible from run to run. Consequently, the effect of velocity on the rate of evaporation of a test chemical is expected to be uniform from run to run. Steady-state concentrations can be attained in less than 1 hour at an air exchange rate of about 5 per hour.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.