Abstract
Temporary parent vessel occlusion performed to establish a high-flow interpositional bypass carries the risk of infarcts. The authors investigated the feasibility of a novel technique to establish a high-flow bypass without temporary parent vessel occlusion in order to lower the risk of ischemic complications. In 10 New Zealand white rabbits, a carotid artery side-to-end anastomosis was performed under parent artery patency with a novel endovascular balloon device. Intraoperative angiography, postoperative neurological assessments, and postoperative MRI/MRA were performed to evaluate the feasibility and safety of the novel technique. A patent anastomosis was established in 10 of 10 animals; 3 procedure-related complications occurred. No postoperative focal neurological deficits were observed. The MRI/MRA findings include no infarcts and bypass patency in 50% of the animals. The authors demonstrated the feasibility of an endovascular assisted, nonocclusive high-flow bypass. Future refinement of the device and technique in an animal model is necessary to lower the complication rate and increase patency rates.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.