Abstract

Prospective motion correction using data from optical tracking systems has been previously shown to reduce motion artifacts in MR imaging of the head. We evaluate a novel optical embedded tracking system. The home-built optical embedded tracking system performs image processing within a 7 T scanner bore, enabling high speed tracking. Corrected and uncorrected in vivo MR volumes are acquired interleaved using a modified 3D FLASH sequence, and their image quality is assessed and compared. The latency between motion and correction of the slice position was measured to be (19 ± 5) ms, and the tracking noise has a standard deviation no greater than 10 μm/0.005° during conventional MR scanning. Prospective motion correction improved the edge strength by 16 % on average, even though the volunteers were asked to remain motionless during the acquisitions. Using a novel method for validating the effectiveness of in vivo prospective motion correction, we have demonstrated that prospective motion correction using motion data from the embedded tracking system considerably improved image quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.