Abstract

The Boltzmann constant k was measured by comparing the Johnson noise of a resistor at the triple point of water with a quantum-based voltage reference signal generated with a superconducting Josephson-junction waveform synthesizer. The measured value of k = 1.380 651(17) × 10−23 J K−1 is consistent with the current CODATA value and the combined uncertainties. This is our first measurement of k with this electronic technique, and the first noise-thermometry measurement to achieve a relative combined uncertainty of 12 parts in 106. We describe the most recent improvements to our Johnson-noise thermometer that enabled the statistical uncertainty contribution to be reduced to seven parts in 106, as well as the further reduction of spurious systematic errors and electromagnetic interference effects. The uncertainty budget for this measurement is discussed in detail.