Abstract

AbstractAn electrochemical sensing platform based on composite material, consisting of molecularly imprinted polymer coated on graphene oxide (MIP‐GO), was developed for selective and sensitive analysis of amoxicillin (AMOX). The MIP‐GO composite, which was fabricated by sol‐gel polymerization after removal of template molecule, was deposited as a thin film on glassy carbon electrode, and then was electrochemically characterized by cyclic voltammetry and differential pulse voltammetry. The linear response for the determination of AMOX was obtained in the concentration range from 5.0×10−10 to 9.1×10−7 M under the most proper conditions and the detection limit was found to be 2.94×10−10 M.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.