Abstract

A radial electromagnetic cavity has been designed and optimized for the in vivo whole-body exposure of mice to 900-MHz RF fields. Parallel circular plates shorted around the perimeter form the cavity, which is fed at the center in order to excite a cylindrical TEM wave. Plastic housings allow the insertion and equidistant positioning from the exciter of 40 mice, with the electric field parallel to the body axis. The resulting exposure system is highly efficient, featuring more than 80% of the incident power dissipated in the mice. The whole-body average SAR can be determined with remarkable precision by means of straightforward power balance since the RF power leakage from the cavity is extremely low. Fairly uniform exposure of the mice, individually and collectively, has been achieved by means of the symmetric arrangement. This exposure system has been adopted in a replication study on transgenic mice currently being carried out in South Australia, and is being considered for upcoming animal studies in Europe.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.