Abstract

Abstract The existence of Pb which formed polyatomic ions such as PbCl+, PbAr+ would interfere the determination of ultra-trace Am by ICP-MS. An extraction chromatography method for the removal of Pb interference was provided in this work to improve the determination accuracy of ultra-trace Am by ICP-MS. In this study, Sr resin was applied to separate Am and Pb because of the great difference between Am and Pb on adsorption ability. The variable parameters including nitric acid concentration, loading volume, column height and loading amount were investigated to optimize the separation conditions for high recovery of Am (R(Am)) as well as high decontamination factor of Pb (DF(Pb)). The optimal separation method was recommended in the study with the recovery of Am over 99% and the decontamination factor of Pb over 2 × 105. Moreover, the method was successfully applied to the analysis of 241Am in the simulated samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.