Abstract

Traditional capillary isoelectric focusing (cIEF), liquid chromatography (LC) and capillary zone electrophoresis (CZE) still suffered from low resolution for hemoglobinopathy screening. Herein, a 30-mm pH 5.2–7.8 microcolumn IEF (mIEF) array chip was developed for hemoglobinopathy screening. As a proof of concept, adult beta-thalassemia was chosen as a model disease. In the method, blood samples were hemolyzed via hemolysin solution and loaded into the microcolumn. The experiments showed that (i) the species of Hb A, F, A2 and variants were clearly separated in the chip, and the resolution was greatly higher than the ones of LC/CZE/cIEF; (ii) up to 24 samples could be simultaneously analyzed in 12-min run; (iii) the intraday and interday RSDs were respectively 3.32–4.91 % and 4.07–5.33 %. The assays of mIEF to total 634 samples were compared with the ones of LC (n = 327) and PCR (n = 307). The cutoff of 3.5 % HbA2 led to the sensitivity of 100 % and specificity of 89.1 % for the mIEF-based screening; and there was 96.7 % coincidence between the methods of mIEF and PCR if refer Hb A2 and F. The method had the merits of facility, efficiency, specificity and sensitivity in contrast to the currently-used methods, implying its potential to screening of beta-thalassemia and hemoglobinopathies.