Abstract

A flow-injection system with on-line ion-exchange preconcentration on dual columns is described for the determination of trace amounts of heavy metals at μg l −1 and sub-μg l −1 levels by flame atomic absorption spectrometry. The degree of preconcentration ranges from 50- to 105-fold for different elements at a sampling frequency of 60 s h −1. The detection limits for Cu, Zn, Pb and Cd are 0.07, 0.03, 0.5, and 0.05 μg l −1, respectively. Relative standard deviations were 1.2–3.2% at μg l −1 levels. The behaviour of the different chelating exchangers used was studied with respect to their preconcentration characteristics, with special emphasis on interferences encountered in the analysis of sea water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.