Abstract
Implantation of a transvenous endocardial pacemaker is contraindicated in patients with a mechanical tricuspid valve. An epicardial left ventricular pacemaker lead was placed by a transvenous route through the coronary sinus into the lateral cardiac vein in a 58-year-old woman with mechanical aortic, mitral, and tricuspid valves, for permanent pacing due to chronic atrial fibrillation with a slow ventricular rate accompanied by syncope. This lead was then connected to a single-chamber pacemaker. The patient had no problem in the following three months. Placement of an epicardial pacing lead through the coronary sinus provides a safe and effective pacing in patients with a mechanical tricuspid valve, thus obviating major cardiac surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Turk Kardiyoloji Dernegi Arsivi-Archives of the Turkish Society of Cardiology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.