Abstract

An axially symmetric spectrometer is described which evolved from a program of measuring Compton profiles with unusually high geometric efficiency. When fitted with a large-volume Ge detector for combined X-ray and γ-ray spectrometry, such as the 51 mm diameter LO-AX detector from EG&G, it allows Compton profiles to be measured at counting rates in excess of 20000 cps. The axially symmetric configuration is also suited to high-efficiency analyses of thick targets by both XRF and Rayleigh/Compton R/C ratios. The same spectra permit a competitive binary-system analysis based on the shape of the Compton profile. Both this new analysis, which we call DuMond spectrometry, and R/C analysis are applied to studies of osteoporosis in the calcaneus with promising results. The combination of high intrinsic and geometric detection efficiency makes it practical to use very weak sources ( ∼ 100 MBq) and unusually low, localized doses ( ∼ 1 μGy) per reading.