Abstract

For direct and non-sampling determination of the component concentration of a sample housed in a glass container, an axially slanted illumination (ASI) back-scattering Raman scheme that reduces glass background interference has been demonstrated. The strategy was to increase the distance between the spots illuminated by the laser on the glass container and the housed sample in back-scattering measurement. For realization, the laser initially illuminated at a slant through the upper side of the vial wall (sample-unoccupied space) and reach the top of the sample. By this way, fewer number of generated glass photons could be recognized by a detector since they are farther from the focal plane (sample-illumination spot). The concentration of rosuvastatin (2.98–4.14 wt%) in rosulord samples (mixed with five other excipients) was determined using the ASI back-scattering measurement. When the angle of illumination to the vertical axis was 30° and the distance from the center of the laser spot on the glass wall to the center of spot on the sample (DG-S) was 14.9 mm, the sample peaks became more apparent and characteristic due to the reduced glass background. The accuracy of the concentration measurement was superior to that obtained through conventional back-scattering, in which the DG-S was nearly zero. The proposed scheme provides a simple optical setting to suppress the glass background and takes advantage of the sensitivity of Raman analysis through back-scattering measurement, indicating it as an attractive option for through-container analysis.