Abstract

This paper reports on the development of an autonomous monitoring system, capable of continuously measuring the pressure inside the bladder. The capsule features wireless bi-directional communication and can be inductively powered anywhere inside the bladder. Short-circuiting the resonant LC tank for the data transmission maximizes the operating range of the passive telemetry. A novel clock extraction method is presented, based on lock-in of a Schmitt-trigger oscillator into the external radio-frequency field. A prototype was built using only discrete components. The system measures the pressure with a resolution of 0.04 kPa at a sampling rate of 10 Hz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.