Abstract

A novel simple fully automatic on-line magnetic stirring-assisted liquid–liquid microextraction method, based on the lab-in-syringe (LIS) concept, has been developed as an alternative approach for sample pretreatment and atomic spectrometric assays. The analytical process includes the in-syringe reaction of the metal ion with the chelating reagent, the analyte micro-extraction and the subsequent transportation of the extractant to the detection system for electrothermal atomic absorption spectrometric (ETAAS) quantification. This novel platform has been demonstrated for trace silver determination in various types of water samples. The method is linear from 19 to 450ngL−1 using a small volume of extraction solvent of 120μL. The entire procedure is accomplished within 240s resulting in a sampling frequency of 15h−1. The enhancement factor is 80, while the detection limit and the precision are 5.7ngL−1 and 3.3%, respectively. The developed method was evaluated by analyzing standard reference materials and spiked water samples with satisfactory recoveries.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.