Abstract
In this article we describe an attenuator device for reducing/controlling the pulse detector counting losses at a high counting rate. The electronics are based on a direct measure of the detector dead time from the analog output signal at the end of the detection chain. Taking into account this parameter the attenuator device decides to reduce/enhance the number of photons that arrive at the detector by inserting/extracting the necessary number of attenuation foils in the x-ray beam path. In that way the number of events in the incoming signal are reduced and the “apparent dynamic range” of the detector is increased.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.