Abstract

A study, where uranium was online enriched and laser diode spectrometry was determined, was performed with some bottled mineral water samples. A uranium-imprinted polymer (PMDU) was synthesized and characterized for preparation of a specific adsorbent of UO22+ cations. The homemade system preconcentrated uranium in PMDU resin and determined using an Arsenazo III complex at 650 nm in a combined laser diode spectrometer. All analytical parameters of the system were optimized at 5.5 of the retention pH, 6.0 N HClO4 of eluent concentration, 0.05% of Arsenazo III complex, and a 40 mm coil length. The effect of interferent ions was also investigated and LOD and LOQ values were found to be 0.54 and 1.80 ng mL-1 respectively. Sample throughput was 12 h-1, the preconcentration factor was 50, and RSD% value was 1.1. Certified reference materials of TMDA 52.3 and TMDA 62.2 were quantitatively analyzed and the proposed method was successfully applied to the bottled mineral water samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.