Abstract
Abstract The design of an improved sampler for the automatic collection of air samples during atmospheric halogenated tracer experiments is described. In this approach, each sample is collected in a small volume (20 to 150 ml) syringe using a rack and pinion drive powered by a precision stepper motor. The sampling rate is controlled via a quartz crystal oscillator and clock to yield averaging periods from 2 to 120 minutes per syringe, and as many as twelve syringe samples can be collected sequentially. Start-up of the sampler is pre-selected using a digital clock and calendar. The sampler is battery-powered, portable and designed for rugged field use. Sample return rates as high as 97% have been achieved in a number of large tracer field programs. Typical results from several of these investigations are also presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.