Abstract

Phenytoin is a commonly used anticonvulsant that is highly protein bound with a narrow therapeutic range. The unbound fraction, free phenytoin (FP), is responsible for pharmacologic effects; therefore, it is essential to measure both FP and total serum phenytoin levels. Historically, the Abbott TDx method has been widely used for the measurement of FP and was the method used in our laboratory. However, the FP TDx assay was recently discontinued by the manufacturer, so we had to develop an alternative methodology. We evaluated the Beckman-Coulter DxC800 based FP method for linearity, analytical sensitivity, and precision. The analytical measurement range of the method was 0.41 to 5.30 microg/mL. Within-run and between-run precision studies yielded CVs of 3.8% and 5.5%, respectively. The method compared favorably with the TDx method, yielding the following regression equation: DxC800 = 0.9**TDx + 0.10; r2 = 0.97 (n = 97). The new FP assay appears to be an acceptable alternative to the TDx method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.