Abstract

We developed an automated purge and trap unit connected to a gas chromatograph-mass spectrometer for shipboard determination of unstable volatile organic compounds in seawater. The device used a small column for the rapid desorption of adsorbed compounds, thus eliminating the need for post-desorption cryofocusing. The repeatability (relative standard deviation, RSD; n = 7) was typically < 5%. The detection limits were 0.1–4.3 pM for chloromethane, bromomethane, dichloromethane, iodomethane, dimethyl sulfide, iodoethane, isoprene, bromochloromethane, chloroform, tetrachloromethane, dibromomethane, bromodichloromethane, iodopropane, chloroiodomethane, dimethyl disulfide, dibromochloromethane, bromoform, and diiodomethane. To investigate the stability of seawater samples, we obtained a concentration-time profile of volatile organic compounds using this method during the incubation of a seawater sample with and without the addition of HgCl2 in the dark at 4°C. We found shipboard determination to be suitable and essential for the determination of unstable compounds such as dimethyl sulfide in seawater, as the concentration of dimethyl sulfide increased considerably during the incubation of a seawater sample both with and without the addition of HgCl2. This method permitted the assessment of numerous naturally produced volatile organic compounds that are considered to be important for the chemistry of seawater/atmosphere exchange in the ocean.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.