Building on the spectrophotometric procedure of Yao and Byrne (1998), an automated analysis system has been developed for laboratory and shipboard measurements of total alkalinity at a rate of 6 samples per hour. The system is based on single-point hydrochloric acid (HCl) titrations of seawater samples of a known volume with bromocresol purple as an indicator. The titration is continuously monitored using visible spectroscopy to guide the titration rate according to the real-time pH of the samples. Each titration is terminated just below the equivalence point to achieve a precision and accuracy near 1μmolkg−1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.