Abstract

A fully automated laboratory-based measurement system for characterization of coil system parameters is presented. This method uses an inexpensive personal computer (PC)-controlled stepper motor positioning system in conjunction with a network/spectrum analyzer and an analog-to-digital converter (A/D) board that allows high resolution data acquisition in an unattended manner. A graphical interface was created for complete control of stepper motor movement, measurement, and data acquisition. The system is capable of performing a wide range of measurements that can, either individually or combined, characterize radiofrequency (RF) and gradient coils used in MRI. Measurement methods, theory, and results for conductor and shield current distributions, mutual impedance, and magnetic fields are given. Comparisons with theoretical calculations are included to validate the accuracy and utility of the system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.