Abstract
Structural parameters of the common carotid artery (CCA) have shown to correlate with the risk of cardiovascular disease, but their precise measurement is challenging. We developed an automatic detection system with manual interaction capabilities that can reliably analyze B-mode ultrasound sequences of the CCA over several heart cycles. We evaluated 3824 frames from 40 sequences in two data qualities. Two readers measured the intima media thickness (IMT) and the lumen diameter at two evaluation times (T1/T2). A Bland-Altman analysis of the average IMT showed a bias ± SD of 0.002 ± 0.010 mm (T1), −0.004 ± 0.008 mm (T2) for completely automatic detections and −0.004 ± 0.010 mm (T1), −0.003 ± 0.010 mm (T2) for clips with manual corrections. The combination of automated analysis and manual intervention provides precise parameters as biomarkers for the atherosclerotic process and makes the system suitable for large scale epidemiological research, diagnostic and clinical practice.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.