Abstract
An automated gravimetric titration system for uranium has been developed, evaluated, qualified for analysis in the laboratory's Quality Assurance Program, and placed in daily operation for the analysis of uranium-bearing safeguards samples at the New Brunswick Laboratory (NBL). The system, a direct application of the NBL Modified Titrimetric Method, is capable of nearly doubling the output of an analyst, while maintaining a mean percent error and relative standard deviation of < 0.10% over the range of 15 to 150 mg of uranium. The system is capable of analyzing all sample materials analyzed by manual titration at NBL with comparable degrees of accuracy and precision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.