Abstract
A fully automated analyzer for methylated L-arginine metabolites [N,N-dimethyl-L-arginine (ADMA), N-methylarginine (NMMA) and N,N'-dimethyl-L-arginine (SDMA)] by high-performance liquid chromatography with post-column fluorescence derivatization was developed. This system consists of an on-line extraction, a separation on a reversed phase ion-pair chromatograph, a post-column derivatization by o-phthaladehyde (OPA) and thiol reaction, and fluorescence detection. NMMA, ADMA and SDMA were separated in 40 min with isocratic elution by a combination of octanoate and cyclohexane carboxylate as ion-pair reagents. The eluate was monitored at 450 nm with excitation at 337 nm. The calibration curves for NMMA, ADMA and SDMA showed linearity over the range from 0.05 micromol l(-1) (0.5 pmol on column) to 5.0 micromol l(-1) (50 pmol on column). This method does not require any time-consuming pre-treatment and requires only 10 microl of plasma sample for assay.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.