Abstract

We present a digital autocollimator which will be used in the alignment system of a prototype medium-sized telescope, which is part of the U.S. contribution to the Cherenkov Telescope Array (CTA). The Schwarzschild-Couder optics in the prototype telescope (which is currently under construction) requires the precise alignment of three components: the primary and secondary segmented mirrors, and the gamma-ray camera. The approximately 9 meter separation between the mirrors necessitates remote optical measurement. Our autocollimator will measure the angle of a segment in one mirror relative to the center of the other mirror with a precision better than 5 arcsec over a range of ±0.126°. We present a detailed description of the instrument and describe its performance in the laboratory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.