Abstract

We performed an audit of the insertion of the Single Use Portex Laryngeal Mask in 400 patients. Insertion was successful at first attempt in 335 out of 400 patients (83.8%). However in 15 patients (3.8%), the Portex laryngeal mask could not be placed despite repeated attempts. In 12 of these 15 patients (80%), a standard Laryngeal Mask Airway (LMA) was successfully placed. After the completion of the audit, 22 out of 29 anaesthetists (75.9%), who had inserted > or = 5 Portex laryngeal masks, considered it inferior to the standard LMA. It would appear to us that the Portex laryngeal mask might need some design modifications to be a real alternative to the standard LMA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.