Abstract

Supersonic shear wave elastography is a novel ultrasound technology that allows objective evaluation of soft tissue stiffness. The purpose of this study is to report an artifact in supersonic shear wave elastography that may strongly affect the accuracy and quality of the measurement of soft tissue stiffness, and to step-by-step describe how the artifact was found and how it was verified by a series of experiments under different conditions. There were a total of three longitudinal band-like artifacts, 1.3 cm apart from each other on the entire image along the length of the SuperLinear SL15-4 transducer surface, with the middle one at the center of the image. When using supersonic shear wave elastography, users should be keenly aware of the possible occurrence of the artifact and avoid it by choosing the quantification measurement regions of interest (ROIs) away from the artifact regions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.