Abstract
Abstract In this paper, a novel four degrees-of-freedom (4DOF) articulated parallel forceps mechanism with a large orientation workspace (±90deg in pitch and yaw, 360deg in roll rotations) is presented for robotic minimally invasive surgery. The proposed 3RSR-1UUP parallel mechanism utilizes a UUP center leg that can convert thrust motion of the 3RSR mechanism into gripping motion. This design eliminates the need for an additional gripper actuator, but also introduces the problem of unintentional gripper opening/closing due to parasitic motion of the 3RSR mechanism. Here, position kinematics of the proposed mechanism, including the workspace, is analyzed in detail, and a solution to the parasitic motion problem is provided. Human-in-the-loop simulations with a haptic interface are also performed to confirm the feasibility of the proposed design.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.