An array of extracellular microelectrodes containing 25 recording and 6 stimulating electrodes was fabricated using microelectronics technology. Ventricular myocardial cells from 8- to 10-day chick embryos were cultured on the surface of the microelectrode array. Extracellular potentials were recorded simultaneously from multiple sites. Simultaneous recordings of extracellular and transmembrane potentials were made from single sites. Extracellular potentials were also recorded simultaneously with cell motion from single sites. Cells were paced by means of electrical stimuli applied via the stimulating electrodes. Conduction velocity in a strip of cells varied linearly as a function of temperature from 0.21 m/s at 26 degrees C to 0.38 m/s at 36.5 degrees C.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.