Abstract

A unique detector design incorporating a 6Li-based capture medium, ZnS scintillator, and wavelength shifting optical fibers is the basis of a new neutron coincidence counter for measurements of plutonium in highly impure residues. The sensor elements have a high efficiency for detecting neutrons and exhibit excellent gamma-ray discrimination based on pulse-shape analysis. The short die-away time of the counter that is based on these detector elements allows coincidence-gate settings shorter than 10μs. This qualifies the technology for measurements of materials with high yields of uncorrelated neutrons from 241Am(α,n) reactions. The characteristics of the new neutron counter will be illustrated with test data from measurements of plutonium, 252Cf, and gamma-ray sources. The integrated electronics design of the new detector also permits the simultaneous but independent measurement of both neutrons and gamma rays. Recent test results that illustrate some unique applications of the sensor's versatility will also be presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.