Abstract
The design of a concentric cylinder calorimeter for measuring the apparent thermal conductivity of MLI blankets is presented. Unlike similar devices where a liquid cryogen is used to control the cold boundary temperature and the cryogen boil-off rate is used to obtain the heat transfer through blanket, the design presented in this paper utilizes mechanical refrigerators to control the boundary temperatures and a heat rate meter to determine the heat load. This approach ensures two unique features of the apparatus. First, the use of cryocoolers enables the user to set the boundary temperatures anywhere within the operating range of the refrigerators and therefore permits a wide range of temperature and temperature differences with the measurement. The other unique feature is that the total heat transfer through the blanket is obtained by measuring the heat conducted along a cold cylinder support rod of known thermal conductivity. To determine the absolute thermal conductivity, a calibration is needed to eliminate the temperature related effects on the support rod.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.