A device for the preparation of polyacrylamide tube gels for pore gradient gel electrophoresis is described. The glass tubes are arranged in the shape of a hollow cylinder, of which the outer as well as the inner surface can be cooled. By this arrangement, the maximum distance the heat of reaction has to be conducted is 7.5 mm. According to the outer diameter of the tubes, the number of gels held by one device ranges from 125 (5 mm) to 500 (2.5 mm). The gels are demonstrated to be identical, and allow for highly reproducible separations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.