While considering the importance of the interface between amputees and prosthesis sockets, we study an anthropomorphic prosthesis socket whose size can be dynamically changed according to the requirements of the residual limb. First, we introduce the structure and function of the anthropomorphic prosthesis socket. Second, we study the dynamic model of the prosthesis system and analyze the dynamic characteristics of the prosthesis socket system, the inputs of an oscillometric pump, and the control mechanism of force-sensitive resistor (FSR) pressure signals. Experiments on 10 healthy subjects using the designed system yield an average detection result between 102 and 112 kPa for the FSR pressure sensor and 39 and 41 kPa for the oscillometric pump. Results show the function of the FSR pressure signal in maintaining the contact pressure between the sockets and the residual limb. The potential development of an auto-adjusted socket that uses an oscillometric pump system will provide prosthetic sockets with controllable contact pressure at the residual limb. Moreover, this development is an attractive research area for researchers involved in rehabilitation engineering, prosthetics, and orthotics.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.