Abstract
In conventional intermittent hemodialysis, the overall mass transfer coefficient ( K o) of a dialyser is mostly calculated at zero ultrafiltration and at relatively high dialysate flow rates. In continuous arterio-venous hemodiafiltration (CAVHD), the dialysate flow rates are low as comparable to the rates of ultrafiltration flows, making the dialysis treatment as slow as possible. Therefore the overall mass transfer coefficient ( K d) of a CAVHD hemofilter has to be calculated in the presence of ultrafiltration. A mathematical model of CAVHD is presented in order to calculate the diffusive mass transfer coefficient ( K d) for a solute when blood, filtrate and dialysate flow rates and solute concentrations are known. The ultrafiltration volume flux ( J v) is assumed to vary linearly along the axial direction of the hemofilter. The calculated mass transfer coefficient K d shows that at high values of dialysate flow and low values of ultrafiltration, the overall mass transfer coefficient ( K d) of a CAVHD hemofilter equals mass transfer coefficient ( K o) of a dialyser in conventional intermittent hemodialysis. Also, the calculated mass transfer coefficient K d shows no significant differences when the ultrafiltration volume flux is assumed to be constant along the length of the hemofilter if no backfiltration occurs in the hemofilter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.