Abstract

AbstractWe consider the design of asymmetric gradient coils in a conventional cylindrical bore magnetic resonance imaging system. The gradient coils are switched on and off during the scan, and are therefore subject to Lorentz forces that cause them to buckle during the switching process. This in turn creates a pressure wave within the coil, and that gives rise to acoustic noise. We present a simplified linearized model for the deflection of the coil due to electromagnetic forces, which is amenable to solution using analytical methods. Closed‐form solutions for the coil deflection and the pressure pulse and noise level within the coil are obtained. These are used to design new coil winding patterns so as to reduce the acoustic noise. Sample results are shown both for unshielded and shielded gradient coils. Extensions of this model are indicated, although it is suggested that the advantages of the present closed‐form solutions might then not be available, and fully numerical solutions may be needed instead. © 2007 Wiley Periodicals, Inc. Concepts Magn Reson Part B (Magn Reson Engineering) 31B: 218–236, 2007

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.